Data of shale samples from the Dawuba Formation

Listed in ScienceDB

Total Organic Carbon (TOC) AnalysisTotal organic carbon (TOC) analysis was performed on the shale samples.

Description

The analysis was carried out using a LECO CS230 carbon‑sulfur analyzer (LECO Corporation, USA). The samples were first crushed and ground to 0.075–0.18 mm, and approximately 0.1 g of each sample was accurately weighed into a crucible.

An excess of dilute hydrochloric acid (10% HCl) was added to remove inorganic carbon. After the reaction was complete, the residue was washed with deionized water until neutral and then dried at 60 °C. About 1 g of iron chip flux and 1 g of tungsten granule flux were added to the dried crucible containing the sample, and the crucible was then analyzed by the instrument. The instrumental working conditions were: combustion temperature 1350 °C, carrier gas high‑purity oxygen at a flow rate of 3 L/min.

Read the rest (5 more)

Instrument reference standards (501‑506 and 501‑679) were inserted after every 20 samples. The analytical error was controlled within ±5%.Major Element AnalysisMajor element analysis was carried out on the shale samples. The analysis was performed with an inductively coupled plasma optical emission spectrometer (ICP‑OES, PerkinElmer Optima 8300V).

The shale samples were crushed to pass 300 mesh (<50 μm) and dried at a temperature of 24 °C and a relative humidity of 55%. An accurately weighed portion of about 0.05 g of sample was placed in a PTFE digestion vessel, and a mixed acid (HF:HNO₃:HClO₄ = 10:5:3) was added. Digestion was performed at 180 °C for approximately 12 hours.

After complete digestion, the solution was made up to 50 mL with 1% HNO₃. The measurement conditions were: RF power 1200 W, plasma gas (Ar) flow 15 L/min, auxiliary gas (Ar) flow 0.2 L/min, nebulizer gas (Ar) flow 0.8 L/min, peristaltic pump speed 1.5 mL/min, and each measurement was repeated three times. The analytical error was controlled within ±5%.Trace and Rare Earth Element AnalysisTrace and rare earth element analysis was performed on the shale samples.

The analysis was carried out using an inductively coupled plasma mass spectrometer (ICP‑MS, Agilent 7700 series). The sample pretreatment was the same as that used for major element analysis: the sample solution was prepared by mixed‑acid digestion (HF:HNO₃:HClO₄ = 10:5:3). The measurement conditions were: RF power 1300 W, carrier gas (Ar) flow 0.9 L/min, auxiliary gas (Ar) flow 1.2 L/min, plasma gas (Ar) flow 15 L/min, lens voltage 6 V, pulse stage voltage 850 V, quadrupole offset 0, detector chamber vacuum 1×10⁻⁶ Pa, and sample uptake rate 1.0 mL/min.

The analytical error was controlled within ±5%.

Links

Where it is published

Catalogue records · 1

Topics

Inferred from text
Geochemistry 72%
Provenance · 1 source records, 11 field assertions
SourceKeyLast seenRaw
ScienceDB10.57760/sciencedb.3758312 d agoJSON v1
FieldAssertionExtractorEvidence
access_levelsource · scidb cnconnector:scidb_cn@1.0.0
concepts[field].anzsrc:group:3703enrichment · scidb cntaxonomy-embedding@1.0.0title+keywords+description (72%)
concepts[field].local:field:earth-environmentalmapping · scidb cnconnector:scidb_cn@1.0.0
concepts[field].local:field:engineeringmapping · scidb cnconnector:scidb_cn@1.0.0
concepts[field].local:field:humanitiesmapping · scidb cnconnector:scidb_cn@1.0.0
concepts[field].local:field:life-sciencesmapping · scidb cnconnector:scidb_cn@1.0.0
concepts[field].local:field:social-sciencemapping · scidb cnconnector:scidb_cn@1.0.0
descriptionsource · scidb cnconnector:scidb_cn@1.0.0/metadata/dc/description
licensesource · scidb cnconnector:scidb_cn@1.0.0/metadata/dc/rights
publication_datesource · scidb cnconnector:scidb_cn@1.0.0
titlesource · scidb cnconnector:scidb_cn@1.0.0/metadata/dc/title