Polarization calibration of the Helioseismic and Magnetic Imager (HMI) onboard the Solar Dynamics Observatory (SDO)

Listed in National Center for Atmospheric Research

As part of the overall ground-based calibration of the Helioseismic and Magnetic Imager (HMI) instrument an extensive set of polarimetric calibrations were performed.

Description

This paper describes the polarimetric design of the instrument, the test setup, the polarimetric model, the tests performed, and some results. It is demonstrated that HMI achieves an accuracy of 1% or better on the crosstalks between Q, U, and V and that our model can reproduce the intensities in our calibration sequences to about 0.4%.

The amount of depolarization is negligible when the instrument is operated as intended which, combined with the flexibility of the polarimeter design, means that the polarimetric efficiency is excellent.

Links

Get the data

Where it is published

Catalogue records · 1

Topics

Provenance · 1 source records, 7 field assertions
SourceKeyLast seenRaw
National Center for Atmospheric Researcha3d0a1f6-486d-4738-b0a6-d15e236aaada8 d agoJSON v1
FieldAssertionExtractorEvidence
concepts[field].local:field:earth-environmentalmapping · data ucar educonnector:data_ucar_edu@1.0.0
concepts[field].local:field:ocean-atmosphericmapping · data ucar educonnector:data_ucar_edu@1.0.0
created_datesource · data ucar educonnector:data_ucar_edu@1.0.0
descriptionsource · data ucar educonnector:data_ucar_edu@1.0.0/notes
publication_datesource · data ucar educonnector:data_ucar_edu@1.0.0
titlesource · data ucar educonnector:data_ucar_edu@1.0.0/title
updated_datesource · data ucar educonnector:data_ucar_edu@1.0.0