Constarium
← Search

Data · dataset · 2022

e-Celsius P022P Capsule Validity Reliability Study

Listed in Borealis

The purpose of this data set was to determine the validity and reliability of the P022-P e-Celsius core temperature capsules which has not been validated previously or assessed for bias and accuracy relative to manufacturer specifications.

Description

We used a ciculating water bath and reference thermometer to test 24 capsules at 7 temperature plateaus between 35C and 42C for 10 samples at each plateau.

Links

Where it is published

Catalogue records · 1

Topics

Provenance · 1 source records, 9 field assertions
SourceKeyLast seenRaw
Borealisdoi:10.5683/SP3/CHUSIB10 d agoJSON v1
FieldAssertionExtractorEvidence
concepts[field].dataverse_subject:medicine-health-and-life-sciencessource · borealisdata caconnector:borealisdata_ca@1.0.0/subjects
concepts[field].local:field:life-sciencesmapping · borealisdata caconnector:borealisdata_ca@1.0.0/subjects
concepts[field].local:field:medicine-healthmapping · borealisdata caconnector:borealisdata_ca@1.0.0/subjects
created_datesource · borealisdata caconnector:borealisdata_ca@1.0.0
descriptionsource · borealisdata caconnector:borealisdata_ca@1.0.0/description
publication_datesource · borealisdata caconnector:borealisdata_ca@1.0.0
titlesource · borealisdata caconnector:borealisdata_ca@1.0.0/name
updated_datesource · borealisdata caconnector:borealisdata_ca@1.0.0
version_labelsource · borealisdata caconnector:borealisdata_ca@1.0.0