Interference of Hb D Variant on the Measurement of HbA1c by Lifotronic H8 Ion-Exchange High-Performance Liquid Chromatography (HPLC)

Listed in Teesside University Research Data Repository

Hemoglobin A1c (HbA1c) is used to monitor long-term glycemic control.

Description

In the presence of hemoglobin variants, erroneous results may be observed in HbA1c measurements. In this case, we describe a patient whose hemoglobin A1c (HbA1c) result was measured with an ion-exchange high-performance liquid chromatography (HPLC) device and found to be 3.1%.

Hemoglobin D(HbD) variant was detected in electrophoresis measurements for further investigation, and the patient's true HbA1c value was 5.1%. In case of very low HbA1c results or measurements inconsistent with the patient's clinic, analysis with methods such as electrophoresis resistant to hemoglobin variant interferences is recommended instead of ion exchange HPLC.

Links

Where it is published

Catalogue records · 1

Topics

Provenance · 1 source records, 7 field assertions
SourceKeyLast seenRaw
Teesside University Research Data Repositoryoai:data.mendeley.com/54gnp4xvn5.117 h agoJSON v1
FieldAssertionExtractorEvidence
access_levelsource · researchdata tees ac ukconnector:researchdata_tees_ac_uk@1.0.0
concepts[field].local:field:earth-environmentalmapping · researchdata tees ac ukconnector:researchdata_tees_ac_uk@1.0.0
concepts[field].local:field:medicine-healthmapping · researchdata tees ac ukconnector:researchdata_tees_ac_uk@1.0.0
descriptionsource · researchdata tees ac ukconnector:researchdata_tees_ac_uk@1.0.0/metadata/dc/description
licensesource · researchdata tees ac ukconnector:researchdata_tees_ac_uk@1.0.0/metadata/dc/rights
publication_datesource · researchdata tees ac ukconnector:researchdata_tees_ac_uk@1.0.0
titlesource · researchdata tees ac ukconnector:researchdata_tees_ac_uk@1.0.0/metadata/dc/title