Data · dataset · 2025
Dataset for Fluorescent Nuclear Track Detectors for out-of-field neutron dosimetry in proton therapy
Listed in Teesside University Research Data Repository
This dataset was generated in: Schmidt, S., Muñoz, I. D., Yukihara, E. G., Vedelago, J. (2025).
Description
Fluorescent Nuclear Track Detectors for out-of-field neutron dosimetry in proton therapy. Submitted to Medical Physics.
Dataset overview: This dataset contains fluorescence (“fl”) and reflection (“re”) microscopy images of alumina-based Fluorescent Nuclear Track Detectors (FNTDs). These detectors were used to measure out-of-field neutron dose values in a clinical proton beam. Experimental setup: Phantom: RW3 and PMMA phantom - Dimensions: 30 cm × 30 cm × 30 cm - Target volume: 10 cm × 10 cm × 10 cm, located at 10 cm to 20 cm depth Detector placement - For RW3-I and PMMA-I, all three FNTDs were positioned 8.1 cm downstream of the phantom isocenter - For RW3-II and PMMA-II, all three FNTDs were positioned 14.1 cm downstream of the phantom isocenter - No lateral (x-direction) shift - The FNTDs were placed 1 cm to 2 cm above the isocenter level Irradiation context - Out-of-field neutron dosimetry in primary proton beams Image acquisition: Detector readout was conducted with a dedicated confocal laser scanning microscopy (FXR700RG).
Read the rest (2 more)
Each detector was scanned at multiple x–y positions, resulting in 175 x–y image locations per FNTD. At each x–y location, a z-stack consisting of 11 images was recorded. For every image, two channels were acquired: - fl – fluorescence image (ionizing particle tracks) - re – reflection image (surface and structural information) The structure of each zip file is summarized below: Component Value ------------------------- ----------- Number of detectors 3 x–y images per detector 175 z-stack slices per image 11 Channels per image 2 Configuration file 1 -------------------------------------- A total of 11,550 images per file (~ 5.5 GB unzipped) File naming convention: Each image name follows the structure FNTD_aa_ybb_xcc_zddee_f.png where aa = detector number (01, 02, ...) bb = y index cc = x index dd = z slice index ee = channel ("fl" or "re") f = detector index without leading zero (0, 1, ...) Note 1: To properly visualize 16-bit images, it may be necessary to apply intensity scaling (e.g. using ImageJ/FIJI).
If not, the images might appear black. Note 2: For images with x indices of 1 or 2, the dominant signal is due to tracks from a Cm-244 source, which is used for inter-detector sensitivity correction. Therefore, out-of-field analysis is only conducted for images with different x indices.
Links
Where it is published
- DOI doi.org/10.17632/m7xnggdfjm.1 ↗
DOI / persistent id · from researchdata tees ac uk
Catalogue records · 1
- OAI-PMH record data.mendeley.com/oai?verb=GetRecord&metadataPrefix=oai_dc&identifier=oai%3Adata… ↗
metadata API · from researchdata tees ac uk
Topics
- From keywords
- Earth & Environmental Science · Medicine & Health · Physics
- Inferred from text
- Image 75% · Microscopy 75%
Provenance · 1 source records, 10 field assertions
| Source | Key | Last seen | Raw |
|---|---|---|---|
| Teesside University Research Data Repository | oai:data.mendeley.com/m7xnggdfjm.1 | 16 h ago | JSON v1 |
| Field | Assertion | Extractor | Evidence |
|---|---|---|---|
| access_level | source · researchdata tees ac uk | connector:researchdata_tees_ac_uk@1.0.0 | |
| concepts[field].local:field:earth-environmental | mapping · researchdata tees ac uk | connector:researchdata_tees_ac_uk@1.0.0 | |
| concepts[field].local:field:medicine-health | mapping · researchdata tees ac uk | connector:researchdata_tees_ac_uk@1.0.0 | |
| concepts[field].local:field:physics | mapping · researchdata tees ac uk | connector:researchdata_tees_ac_uk@1.0.0 | |
| concepts[modality].local:modality:image | enrichment · researchdata tees ac uk | keyword-concept-rules@1.0.0 | title+description (75%) |
| concepts[modality].local:modality:microscopy | enrichment · researchdata tees ac uk | keyword-concept-rules@1.0.0 | title+description (75%) |
| description | source · researchdata tees ac uk | connector:researchdata_tees_ac_uk@1.0.0 | /metadata/dc/description |
| license | source · researchdata tees ac uk | connector:researchdata_tees_ac_uk@1.0.0 | /metadata/dc/rights |
| publication_date | source · researchdata tees ac uk | connector:researchdata_tees_ac_uk@1.0.0 | |
| title | source · researchdata tees ac uk | connector:researchdata_tees_ac_uk@1.0.0 | /metadata/dc/title |